2-11
Chapter 2 Network System Construction Wireless Network
CAUTION
When the measurement unit of BP is "kPa", corresponding waveform and numeric data
will not be transmitted. When using a wireless network, use "mmHg" for the BP
measurement unit.
BP waveform with a scale above the set scale can not be properly transmitted. When
transmitting the BP waveform, pay attention to the displayed BP scale.
If the measurement unit of CO
2
concentration is mmHg and [99mmHg] is selected for
"Upper Limit of CO
2
(mmHg) Transmission" ([Initial Settings]>[System]>[Telemeter]), the
CO
2
value of 100mmHg or above will be transmitted as 99mmHg.
When using the Nellcor
TM
unit, the PR value of 255bpm or above will be transmitted to
the central monitor as 254bpm.
REFERENCE
The waveform not displayed on the home display can not be transmitted.