Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual
Chapter 3 LOGIQ Totus Setup 3 - 79
3-7-9-7 Print Button Settings
Use the Print Button Settings page to set parameters for print buttons.
3.
When connected to the network through a wireless LAN, to send images to the DICOM device over the wireless LAN, enable Send Images
via Wireless.
NOTE: If disabled, images spooled in the Spooler will be sent when the Ultrasound system is connected to the wired network.
4.
When connected to the network through a wireless LAN, to receive images from the DICOM device over the wireless LAN, enable Receive
Images Wireless.
NOTE: If disabled, images will be received when the Ultrasound system is connected to the wired network.
5.
Under Send To: Default Service, select the default Send To service.
6.
To allow communication between the Ultrasound system and the Koios Server VM (installed in the Ultrasound system) even when there is no
network, enable Process Dicom Job Without Network.
Table 3-42 To Configure Print Button Settings
Step Corresponding Graphic
1.
Navigate to Connectivity > Connection Manager > Advanced Settings.
2.
Select Print Button Settings.
3.
To allow the pressing and holding of a print key to replace an image, enable Allow press and hold print key to replace an image.
4.
To allow a smart capture area, enable Enable Smart Capture Area.
5.
To add a title bar to the DICOM image, enable Add Titlebar information to Multiframe loops.
6.
To store DICOM multiFrame images when collecting RF Data, enable Store Dicom MultiFrame When Collecting RF Data.
7.
To store a 2D loop with timeline data, enable Store 2D Loop with Timeline Data.
8.
To add scan parameter(s) to the DICOM image, enable Add Scan Parameter information to Multiframe Loops.
9.
To store DICOM MultiFrame images in Quantitative Analysis mode, enable Store Dicom MultiFrame in QAnalysis pack.
10.
Under Patient List Print - Font Size, select the font size.
11.
Under P[1-4] Key Sound, select None, Click, Chimes, Ding, Ding-Dong, or Whoosh.
Table 3-41 To Configure Transfer Settings (Continued)
Step Corresponding Graphic

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ Totus and is the answer not in the manual?

GE LOGIQ Totus Specifications

General IconGeneral
BrandGE
ModelLOGIQ Totus
CategoryMedical Equipment
LanguageEnglish

Related product manuals