Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual
3 - 80 Section 3-7 - Setting Up Connectivity and Tips
3-7-9-8 Image Numbering
Use the Patient Screen page to set parameters for the image numbering.
Table 3-43 To Configure Image Numbering
Step Corresponding Graphic
1.
Navigate to Connectivity > Connection Manager > Advanced Settings.
2.
Select Image Numbering.
3.
To use Scan Assistant image reordering/renumbering, enable Use Scan Assistant image ordering/numbering.
4.
Under Image Numbering Scheme, select the preset that defines how the images are numbered during the examination/scanning process.
5.
To show the image number on the Active Images screen, enable Show image number on Active Images screen.
6.
Under Image Reordering Scheme, select the preset that defines how the images are numbered during image reorder operation.