GE HealthCare CONFIDENTIAL
DIRECTION 5936427, REV. 1 LOGIQ TOTUS Basic Service Manual
Section 8-8 - Replacing Main Console 8 - 139
8-8-11 Replacement of the Air Filter Set-U
This table includes information specific to these instructions. For more information 8-2-6 "Tools needed
for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6
8-8-11-1 Preparations
1.) Remove the T2 Side Cover L Assy as described in 8-8-7 "Replacement of the T2 Side Cover L
Assy" on page 8-135
2.) Remove the T2 Side Cover R Assy as described in 8-8-8 "Replacement of the T2 Side Cover R
Assy" on page 8-136
3.) Remove the Rear Cover Mid Assy-T as described in 8-8-10 "Replacement of the Rear Cover Mid
Assy-T" on page 8-138
8-8-11-2 Removal Procedure
8-8-11-3 Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in Section 8-
12 on page 8-223
Table 8-73 Manpower, Time, Tools, and PPE
Manpower Time Tools PPE
One person Approximately 5 minutes No tools required
Cut resistant gloves
Knee Pad
Step Corresponding graphic
1.
Release 2 snap fits to remove the Unity Air Filter Side
from the T2 Side Cover L/R Assy.
2.
Release 3 snap fits to remove the Unity Air Filter Rear
from the Rear Cover Mid Assy-T.