Section 8-8 - Replacing Main Console 8 - 141
8-8-13 Replacement of the Castor Dual Lock N9
This table includes information specific to these instructions. For more information 8-2-6 "Tools needed
for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6
8-8-13-1 Preparations
1.) Remove the Unity Base Tray as described in 8-8-2 "Replacement of the Unity Base Tray" on page
8-130
2.) Remove the Front Cover Assy-T as described in 8-8-3 "Replacement of the Front Cover Assy-T" on
page 8-131
3.) Remove the T2 Side Cover L Assy as described in 8-8-7 "Replacement of the T2 Side Cover L
Assy" on page 8-135
4.) Remove the T2 Side Cover R Assy as described in 8-8-8 "Replacement of the T2 Side Cover R
Assy" on page 8-136
5.) Remove the Rear Cover Mid Assy-T as described in 8-8-10 "Replacement of the Rear Cover Mid
Assy-T" on page 8-138
6.) Remove the Base Cover Set-T as described in 8-8-12 "Replacement of the Base Cover Set-T" on
page 8-140
8-8-13-2 Removal Procedure
8-8-13-3 Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in Section 8-
12 on page 8-223
Table 8-75 Manpower, Time, Tools, and PPE
Manpower Time Tools PPE
Two people Approximately 15 minutes
24mm Socket or
Adjustable wrench
Cut resistant gloves
Knee Pad
Step Corresponding graphic
1.
Unscrew the nut to remove the Castor Dual Lock
N9 from the console.
NOTICE
Press the caster wheel locks and ensure that they keep the system from moving.
WARNINGWARNING
Proceed with extreme caution and avoid system tipping over. The system weighs
approximately 85kg.