GE HealthCare CONFIDENTIAL
DIRECTION 5936427, REV. 1 LOGIQ TOTUS Basic Service Manual
8 - 150 Section 8-8 - Replacing Main Console
8-8-21 Replacement of the Unity Power Cord Fix Brkt
This table includes information specific to these instructions. For more information 8-2-6 "Tools needed
for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6
8-8-21-1 Removal Procedure
8-8-21-2 Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in Section 8-
12 on page 8-223
Table 8-83 Manpower, Time, Tools, and PPE
Manpower Time Tools PPE
One person Approximately 5 minutes Philips Screwdriver
Cut resistant gloves
Knee Pad
Step Corresponding graphic
1.
Unscrew 2 screws to remove the Unity Power Cord Fix Brkt from
the console.