EasyManuals Logo

GE LOGIQ Totus User Manual

GE LOGIQ Totus
644 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #478 background imageLoading...
Page #478 background image
GE HealthCare CONFIDENTIAL
DIRECTION 5936427, REV. 1 LOGIQ TOTUS Basic Service Manual
8 - 152 Section 8-8 - Replacing Main Console
8-8-22-2 Removal Procedure
8-8-22-3 Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in Section 8-
12 on page 8-223
Steps Corresponding Graphic
1. Note: If UBC-T is installed, remove the UBC-T as
described in
8-11-24 "Replacement of the UBC-
T" on page 8-219
Unscrew 4 screws to remove the Unity Cable Hide
Brkt from the console
2. Push button of the ACDC Output cable’s connector to
disconnect the ACDC output cable and the ACDC
Link Cable to the UBE Cable.
3. Unscrew 4 screws to remove the Unity ACDC Assy
from the console.
4. Pull the ACDC Assy and remove from the console.