GE HealthCare CONFIDENTIAL
DIRECTION 5936427, REV. 1 LOGIQ TOTUS Basic Service Manual
8 - 196 Section 8-11 - Replacing Options
8-11-11 Replacement of the TVTR Probe Holder T1
This table includes information specific to these instructions. For more information 8-2-6 "Tools needed
for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6
8-11-11-1 Removal Procedure
8-11-11-2 Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in Section 8-
12 on page 8-223
Table 8-106 Manpower, Time, Tools, and PPE
Manpower Time Tools PPE
One person Approximately 5 minutes No tools required Cut resistant gloves
Step Corresponding graphic
1.
Release 1 snap fit and pull the TVTR Probe Holder T1
to remove from the console.