GE HealthCare CONFIDENTIAL
DIRECTION 5936427, REV. 1 LOGIQ TOTUS Basic Service Manual
8 - 240 Section 8-12 - Functional Checks after FRU replacement (Debrief Guideline)
8-12-89 Replacement of the UECG Assy
8-12-90 Replacement of the VNAV Option Cable Set-U
8-12-91 Replacement of the 3DG DriveBay 2+ EU and Power Adapter Cable
Functional Check Debrief
Script
4-2-3 "Power ON/Boot Up"
Equipment passed all required checks and is ready for
use.
4-3-8 "B-Mode functional checks"
4-3-10 "System CFM and PWD functional checks"
4-3-14 "PW/CW Doppler mode functional checks"
4-3-20 "Basic measurement functional checks"
4-3-22 "ECG functional checks"
Utility->Applications->Settings->ECG->Show ECG tools.
Connect cables. Press Display ECG tab on Touch Panel.
Verify ECG signal displayed on the monitor.
Plug in the ECG cables with nothing attached so they pick
up noise.
4-3-24 "Volume Navigation with inside probe func-
tional checks"
4-2-4 "Power shut down"
Functional Check Debrief
Script
4-2-3 "Power ON/Boot Up"
Equipment passed all required checks and is ready for
use.
4-3-7 "Probe/connector functional checks"
4-3-8 "B-Mode functional checks"
4-3-10 "System CFM and PWD functional checks"
4-3-20 "Basic measurement functional checks"
4-3-24 "Volume Navigation with inside probe func-
tional checks"
4-2-4 "Power shut down"
Functional Check Debrief
Script
4-2-3 "Power ON/Boot Up"
Equipment passed all required checks and is ready for
use.
4-3-7 "Probe/connector functional checks"
4-3-8 "B-Mode functional checks"
4-3-10 "System CFM and PWD functional checks"
4-3-20 "Basic measurement functional checks"
4-3-24 "Volume Navigation with inside probe func-
tional checks"
4-2-4 "Power shut down"