GE HealthCare CONFIDENTIAL
DIRECTION 5936427, REV. 1 LOGIQ TOTUS Basic Service Manual
8 - 242 Section 8-12 - Functional Checks after FRU replacement (Debrief Guideline)
8-12-96 Replacement of the UBE To Monitor HDMI Cable
8-12-97 Replacement of the UBE To OPIO Touch Panel HDMI Cable
8-12-98 Replacement of the UBE TO OPIO HUB USB 3.0 Cable
Functional Check Debrief
Script
4-2-3 "Power ON/Boot Up"
Equipment passed all required checks and is ready for
use.
4-3-8 "B-Mode functional checks"
4-3-10 "System CFM and PWD functional checks"
10-6-4 "Grounding Continuity"
10-6-6 "Earth Leakage Current Test"
4-2-4 "Power shut down"
Functional Check Debrief
Script
4-2-3 "Power ON/Boot Up"
Equipment passed all required checks and is ready for
use.
4-3-7 "Probe/connector functional checks"
4-3-8 "B-Mode functional checks"
4-3-10 "System CFM and PWD functional checks"
4-3-11 "Color Mode functional checks"
4-3-14 "PW/CW Doppler mode functional checks"
4-3-1 "Control panel adjustment and functional
checks"
4-2-4 "Power shut down"
Functional Check Debrief
Script
4-2-3 "Power ON/Boot Up"
Equipment passed all required checks and is ready for
use.
4-3-7 "Probe/connector functional checks"
4-3-8 "B-Mode functional checks"
4-3-10 "System CFM and PWD functional checks"
4-3-11 "Color Mode functional checks"
4-3-14 "PW/CW Doppler mode functional checks"
4-3-1 "Control panel adjustment and functional
checks"
4-2-4 "Power shut down"