ADVANCED OPERATION
239
Fail Icon:
7. Press More[F6]>Trigger[F1] to toggle between a
free run (internal) trigger and an external
trigger.
Trigger:
Free, Ext
Ext trigger input:
3.3V CMOS
8. Press More[F6]> MAX/MIN HOLD[F2] to toggle
the MAX/MIN hold measurements on/off in
the power meter bar.
The MIN/MAX measurements will be displayed in
the power bar meter in the center of the screen.
MIN HOLD
measurement
MIN HOLD
measurement
Note
The return to the normal Spectrum Mode, turn the
power meter off by pressing
Option
Control
>
Power
Meter[F2]>Power Meter[F1]
and toggle the
power
meter off.
Data Logging Power Meter Measurements
When in Power Meter mode, the spectrum
analyzer is able to log the power meter
measurements over a user-defined time period
at user-defined intervals.
1. Press
Save
to enter the save menu.
2. Press Type[F2] and select Power Meter[F7].