Microlab
®
STAR Service Manual
6-37
6.6 Calibrating Pipetting Arm with Channels
Run the following macro program together with Channel Calibration Tool P/N
173960:
• PIP AUTOADJUSTMENT.MCR
Automatic Adjustment procedure:
Channel Calibration Tool P/N 173960
placed onto deck as asked for by the
program. (Appropriate Adjustment
Program has been started.)
After the successful Adjustment ensure all screws on Pipetting Arm and Pipetting
Channels are tight.
Perform a Check Run.
Cover pipetting arm and Instrument.