Microlab
®
STAR Service Manual
4-11
4.2.3.2 Power / Voltage
Using an Uninterruptible Power Supply (UPS) is highly recommended.
The mains plug is on the left-hand side of the instrument towards the rear.
The fuses for the instrument are situated under the mains power switch.
Plug in the mains cables for the computer and the instrument into the same
electrical outlet.
Attention: Connect only to an earth-grounded outlet
Attention: For safety reasons, ensure that the appropriate Fuse is placed into the
mains power switch before switching on the instrument.
Fuses:
P/N 363012 Delayed action FUSE 3.15 A for 230 V
P/N 363013 Delayed action FUSE 6.3 A for 115 V
See section 2.2 Technical Specifications on page 2-6.
Voltage selector
Fuses