7-33
7.8.1 Part List
P/N Description SPC
173330 TIP EJECTOR A
173332 STOP DISK A
173520 DISC 2 A
182105 REPLACE. PIPETTING HEAD 300ul A
182106 REPLACE. PIPETTING HEAD 1000ul A
254167 O-Ring ID3.6x1.45 A
400602 CYL-SCR M2x5 DIN912 A
400621 CYL-SCR M2.5x12 DIN912 A
409103 DISC M2 DIN125A A
511071 OMNI FIT 15M A
173310 GREASE 100 GR TOPAS AK50 A
7.8.2 Function
The following features are implemented in the Pipetting Head:
Tip coupling with CO-RE technology.
Tip presence is registered.
Capacitive Liquid Level Detection (for Aspiration and Dispensing).
Pressure Liquid Level Detection (for Aspiration only, by using a new disposable
Tip for each aspiration step).
MAD (Monitored Air Displacement), which monitors aspiration.
Aspiration and dispensing of liquid
Tip ejection with CO-RE technology
ML STAR IVD additionally features
TADM (Total Aspiration and Dispense Monitoring), where aspiration and
dispensing are monitored.