STAR Service Manual
8-17
8.4.3 Replacement
See section 7.7 Pipetting Channels on page 7-26.
Before a Pipetting Channel is replaced, print a hardcopy of all Data from the
Channel by using Service Software.
8.4.4 Remounting:
Instructions are basically the reverse of those for replacement. See section 7.7
Pipetting Channels on page 7-26.
With Service Software:
Check Pipetting Channel settings
Check for accurate Firmware
8.4.5 Adjustment
It is absolutely necessary to run the PIP AUTOADJUSTMENT.MCR to generate
new calibration values for the Pipetting Channel.
Attention: Do not restore any Data to Pipetting Channel (if previously copied from
replaced Pipetting Channel). The Calibration values are no longer valid since the
Pipetting Channel is physically replaced by another one.