EasyManua.ls Logo

Hamilton Microlab STAR - Instrument Check Procedure

Hamilton Microlab STAR
284 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Microlab
®
STAR Service Manual
4-27
4.2.6.1 Instrument Check Procedure
Before running any adjustment programs as described in chapter 6 Adjustment and
Calibration on page 6-1, perform an Instrument check procedure with Macro
Program “INSTRUMENT CHECK.MCR”.
606867 Instrument Check.mcr
passed?
606827 PIP Autoadjustment.mcr
606867 Instrument Check.mcr
passed?
606828 PIP XY Manual Adjustment.mcr
Passed?
yes
no
no
improved?
yes
no
yes
606825 Arm Z-Difference .mcr
until passed
606828 PIP XY Manual Adjustment.mcr
606826 Arm X-Difference .mcr
until passed
606828 PIP XY Manual Adjustment.mcr
606827 PIP Autoadjustment.mcr
yes
606867 Instrument Check.mcr
Passed?
yes
no
setup / installation of ML STAR instrument
Perform mechanical positioning adjustment
according to result values
if values are far out of tolerance then mechanical
adjustment with adjustment tools must performed
setup sucessful
continue according to installation qualification
no

Table of Contents

Other manuals for Hamilton Microlab STAR