STAR Service Manual
6-5
6.2 Adjustment Tools and Macro Programs
6.2.1 Tools for Pipetting Arm and Channels
Tool Description
Channel Adjustment Tool P/N 173952 For mechanical Channel Adjustment on
Pipetting Arm.
Must be used for preadjustment
whenever a Channel is replaced in
Order to mount the new channel on the
Pipetting Arm.
Predefines:
Parallelity (alignment) in Z Direction
in functional dependence on X- and
Y-Axis.
Linearity (alignment) of all Channels
in Y-Axis.
Channel Calibration Tool P/N 173960
Reference for Pipetting Arm Alignment
/ Adjustment and Pipetting Channel
Calibration
Together with the Adjustment macro
programs, ensures perfect alignment of
Pipetting Arm and all Channels.
All Pipetting Channels must be
mounted and adjusted with Channel
Adjustment Tool first, then calibration
with Channel Calibration Tool can take
place.
A Pointer (Z Tool P/N 173968) is
included with the Channel Calibration
Tool.
CO-RE Tips for Adjustment P/N
235920 can be used with Macro ‘PIP
ALIGNMENT WITH LLD.MCR’.
Z- Tool P/N 173968