Microlab
®
STAR Service Manual
6-4
Order of adjustment and calibration
606828 PIP XY Manual Adjustment.mcr
passed?
improved?
yes
no
606825 Arm Z-Difference .mcr
until passed
606828 PIP XY Manual Adjustment.mcr
606826 Arm X-Difference .mcr
until passed
606828 PIP XY Manual Adjustment.mcr
606827 PIP Autoadjustment.mcr
yes
606867 Instrument Check.mcr
passed?
no
606825 Arm Z-Difference .mcr
until passed
606826 Arm X-Difference .mcr
until passed
mechanical adjustment with adjustment tools
Deadjusted Channel, or replaced Channel
Perform mechanical positioning adjustment
according to result values
if values are far out of tolerance then mechanical
adjustment with adjustment tools must performed
adjustment sucessful
yes
no