2-14
2.4.1 Service Tool Maintenance
In order to provide best service and maintenance to the components of the
Microlab
STAR instrument, the following maintenance procedures should apply to
these service tools at the specified intervals:
The torque wrench P/N 239646 must be subject to an annual check and calibration.
Calibration may performed by any calibration authority; Hamilton Bonaduz AG offers
a calibration service.
The Adjustment Tools, such as the Channel Adjustment Tool P/N 173952 may be
returned periodically to Hamilton Bonaduz AG where a check service is offered.
Note: Adjustment Tools can not be calibrated. If an Adjustment Tool is found to be
out of range it must be replaced with a new one.
Attention: the Microlab
STAR Special Tools are precision tools and must be
handled with due care to provide the best results in adjustment work on the
Instrument.
2.5 Disposal of ML STAR Instruments
After the life cycle of the instrument has terminated, the Microlab STAR may be
shipped back to Hamilton Bonaduz AG. Otherwise local disposal regulations are to
be observed.