9.5 Half-yearly Maintenance
Action Available Program
1. Decontaminate and clean exposed Surfaces
Deck
Docking Station
Loading trays
25
Instrument Covers
Pipetting Arm housing
Autoload ribbon
26
Spillage Trays
Air filter, replace if necessary.
N/A
Attention:
Do not grease Y-Spindles Pipetting Arm
Do not use any Cleaning liquids other than on Spindles to remove old grease.
Avoid any contact with the CO-RE O-Ring as this may shorten its life span.
2. Uncover Pipetting Arm
Check Cable connections onto Pipetting
Channels.
Check and clean Y- Spindle on Pipetting Arm.
Spindle must not be greased at all.
If needed, e.g. when step losses in Y occurs,
Silicon oil may used to apply a thin layer onto
the Y-Spindle.
N/A
3. On Pipetting Heads:
Replace CO-RE O-Rings
Check Stop Disk(s)
check for any scratches, blocked bore, or visible
damage, and replace if such is the case.
Check Tip ejection Tube(s)
check against contamination and damage.
Ensure it glides smoothly.
For replacement
Procedures refer to
section 7.8 Pipetting
Heads on page 7-32.
25
Microlab
®
-STAR Instrument with Autoload Option only
26
dito