STAR Service Manual
6-21
6.4.1 Pipetting Arm degrees of freedom
Note: the above-mentioned Fixing screws may occur as setscrews with hex nuts and washers.
Pipetting Arm Fixing Screws:
When both Pipetting Arm Alignment
Tools are in place, loosen 8 fixing
screws in the rear for tilting around
Z-Axis.
Note: The Z Axis to tilt around will
then be in the rear.
This aligns the Pipetting Arm
parallel to Y-Axis (the X Coordinate
will then be the same for the front
and the rear of Pipetting Arm.
Front Guide Bar:
Not displayed here. By slightly lifting and
lowering, tilting around X-Axis is enabled.
Note: The X Axis to tilt around will then be
in the rear.