Microlab
®
STAR Service Manual
4-17
4.2.4.3 Accessories Installation (ML STAR IVD)
Accessories CD comes together with Carrier
Put CD into CD-ROM-Drive
start Setup.exe Program File from the CD-ROM and carefully follow the
instructions during installation.
4.2.4.4 Service Software Installation
Generally - see section 3 Service Software on page 3-1.
Attention: Do not leave the Service Software permanently installed on a Customer's
PC as this will enable any user of that PC to "play" with the system - with potentially
damaging results!
4.2.5 Software Presettings
4.2.5.1 Overview
User and Service Software are completely independent of one another. Therefore,
settings such as "Com Port settings" must be made for both.
Basically:
Using the Service Software, you can set the Microlab
STAR Instrument
Configurations on the Master Board located inside the Instrument.
Using the User Software, you can set the Microlab
STAR Instrument
Configurations necessary for programming Methods.
4.2.5.2 Instrument Configuration
Check the Instrument Configuration by using service software and print out
summary.
Note: The Instrument and Software Configurations must match. The User Software,
during Method parsing, compares the settings of Instrument and User Software.