EasyManua.ls Logo

Hamilton Microlab STAR - Verification; Volume Verification; Verification Kits

Hamilton Microlab STAR
284 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Microlab
®
STAR Service Manual
6-47
6.9 Verification
6.9.1 Volume Verification
The Verification Kit provides the means to verify instrument functions against
specific acceptance criteria. The gravimetric verification procedure consists of an
easy-to-follow program called from the user Software.
The Pipetter’s Accuracy and Precision will be verified.
6.9.2 Verification Kit’s
The Verification Kit consists of 3 parts:
A Basekit witch may be used for several Hamilton Instruments next to the ML
STAR Instrument, however it is current depending for Countries with either 115
VAC or 230 VAC.
An Instrument specific “SUPPLEMENT” Kit witch contents ML STAR Instrument
specific items.
A “CONSUMABLE KIT” which contains verification fluid, disposables Tips, etc.
Countries with 230 VAC Countries with 115 VAC
P/N 182501
VFV BASEKIT 230V/50Hz
P/N 182502
VFV BASEKIT 115V/60Hz
P/N 182503
VFV SUPPLEMENT ML-STAR
P/N 182506
VFV CONSUMABLE KIT ML-STAR

Table of Contents