Microlab
®
STAR Service Manual
2-6
2.2 Technical Specifications
Item Value
maximum weight of instrument
with the Autoload option and trays installed
8 Channel Version:
16 Channel Version:
ML STAR Instrument packed for
transportation – total weight including
shipping crate:
145 kg
155 kg
< 200 kg
electrical requirements
maximum power consumption:
Voltage:
Frequency:
Delayed action fuse:
115 V∼:
230 V∼:
600W
115 / 230 V∼ -15 % + 10 %
50 / 60 Hz ± 5 %
6.3 A
3.15 A
Operating temperature range:
15-35 °C
relative humidity 30 – 85 % with
no condensation
Storage temperature range:
0 - 55 °C
relative humidity at a maximum of
95 % with no condensation