8.3.3 Replacement
Remove Left Deck Panel to access Master Board, see section 7.4.2 Deck Panels
on page 7-7.
Before the Master Board is replaced, print a hard copy of all Data from the Board
by using Service Software, and keep on file.
Disconnect all cables and unscrew Master Board.
Screw P/N 400656 5 x
Screw P/N 400632 1 x
INTERFACE METAL PLATE P/N 173614 1 x
WASHER P/N 369026 1 x
ISOLATING TAPE P/N 369061 1 x
8.3.4 Remounting:
Instructions are basically the reverse of those for disassembly.
By using Service Software
Data may restored (if copied from replaced Master PCB).
Check Master settings
Check for correct Firmware
8.3.5 Adjustment
It is highly recommended that you run the autoadjustment (PIP
AUTOADJUSTMENT.MCR) to generate new calibration values for X-Drive
Pipetting Arm.