STAR Service Manual
8-14
8.3.3 Replacement
• Remove Left Deck Panel to access Master Board, see section 7.4.2 Deck Panels
on page 7-7.
• Before the Master Board is replaced, print a hard copy of all Data from the Board
by using Service Software, and keep on file.
• Disconnect all cables and unscrew Master Board.
Screw P/N 400656 5 x
Screw P/N 400632 1 x
INTERFACE METAL PLATE P/N 173614 1 x
WASHER P/N 369026 1 x
ISOLATING TAPE P/N 369061 1 x
8.3.4 Remounting:
• Instructions are basically the reverse of those for disassembly.
• By using Service Software
• Data may restored (if copied from replaced Master PCB).
• Check Master settings
• Check for correct Firmware
8.3.5 Adjustment
• It is highly recommended that you run the autoadjustment (PIP
AUTOADJUSTMENT.MCR) to generate new calibration values for X-Drive
Pipetting Arm.