Microlab
®
STAR Service Manual
9-6
Half-yearly Maintenance (continued)
Action Available Program
4. On Pipetting Heads clean and only slightly grease:
• Squeezer Spindle only with brass nuts
• Dispensing Spindle
Service Software:
Single Commands
or
Movements Sensors
5. Clean and slightly grease guide shafts:
• X-Guide Pipetting Arm front and rear
• X-Guide Autoload
27
Service Software:
Single Commands
or
Movements Sensors
6. Remove Deck Panels and check:
• For signs of liquid spillage underneath deck and
inside electronic chamber.
• If necessary: decontaminate and clean surfaces
of deck underneath and electronic chamber
cover.
• Clean Flat Band Cables Groove and ensure
collision free movement of Flat Band Cables.
• All belts, drives and cables for any damage - or
loose screws
• Distance of X-measurement Reader on
Pipetting Arm to magnetic Band
N/A
7. Remove Autoload ribbon
28
and check:
• For signs of liquid spillage underneath.
• If necessary: decontaminate and clean surfaces
of Autoload ribbon underneath.
• Clean Flat Band Cables Groove and ensure
collision free movement of Flat Band Cables.
• All belts, drives and cables for any damage, or
loose screws.
N/A