Microlab
®
STAR Service Manual
9-7
Half-yearly Maintenance (continued)
Action Available Program
8. Recover:
• Autoload ribbon
29
• Deck Panels
• Pipetting Arm
N/A
9. Check and Adjust:
• Pipetting Arm
• Pipetting Channels
• Autoload
30
• Verify instrument with verification kit
• Perform a test run
Adjustment is done
in accordance with
section 6 Adjustment
and Calibration on
page 6-1.
Verification Program
N/A
10.
Check all available Carriers:
• For signs of liquid spillage and damage
• If necessary: decontaminate and clean Carriers.
• Ensure smooth gliding on Deck Panels and
Loading Trays.
• Replace damaged carriers.
N/A
11.
Check Needles if available:
• For signs of damage.
• If necessary: decontaminate and clean Needles
• Replace damaged Needles
N/A
Attention: Items for ML STAR and ML STAR IVD may vary, therefore ensure correct
parts are being used on the Instrument.