Half-yearly Maintenance (continued)
Action Available Program
12. Removal of old files
• Together with the Laboratory Supervisor check
for trace files witch may deleted from the
Operating PC.
• Check for Traces files such as
method / system trace files
communication trace files
• Check for old methods which may be removed
from PC
• Back them up if necessary, else delete
• Check for available disk space onto hard drive,
500 MB minimum are needed
N/A
ML STAR IVD additionally:
13.
• Test tightness of seal for all Pipetting Heads. Maintenance
Procedure according
to ML STAR IVD
Operators Manual
14. • Check the pressure adjustment values of
TADM.
PIP_PRESSURE_SIG
NALS_
CHECK.MCR