STAR Service Manual
7-43
Removal (continued)
View from above:
The Autoload drive is connected with
a total of 6 screws P/N 400632 onto 2
sliding carriages.
4 screws onto the right,
and 2 screws onto the left sliding
carriage.
The belt clamp is mounted with 2
screws P/N 400010 onto the autoload
drive and must be disconnected as
well.
3 screws are involved, however the
middle screw is to remain in place.
Through the hole unscrew:
two screws P/N 400653 holding sheet
metal and removing it, then unscrew
four slide screws P/N 400632
underneath sheet metal.
Sheet metal away from aluminum
block