10-6
Error Handling (continued)
Incident Corrective Action
Flash EPROM, E
2
EPROM
errors (typically after
download of new firmware)
Autoload Drive:
1. Run Macro “SET_EEPROM_I0_DATA.MCR”.
Flash EPROM, E
2
EPROM
errors
Note:
After running of the above mentioned Macros
switch ML STAR Instrument OFF and then ON!
Method on PC runs without
any action on instrument
1. Check Configuration Settings in user Software
and switch from simulation to instrument mode
Configuration mismatch on
starting any method
1. Check the physical Configuration of Microlab
STAR instrument:
Autoload, number of channels, channel type,
etc.
2. Compare physical Configuration with settings
first in Service and then in user Software.