Microlab
®
STAR Service Manual
12-12
Maintenance Checklist (continued):
Remove Deck Panels and
check:
OK REMARKS
For signs of liquid spillage
underneath deck and inside
electronic chamber.
______________________________________
If necessary: decontaminate and
clean surfaces of deck
underneath and electronic
chamber cover.
______________________________________
Clean Flat Band Cables Groove
and ensure collision free
movement of Flat Band Cables.
______________________________________
All belts, drives and cables for
any damage - or loose screws
______________________________________
Distance of X-measurement
Reader on Pipetting Arm to
magnetic Band
______________________________________
Remove Autoload ribbon
35
and check:
For signs of liquid spillage
underneath.
______________________________________
If necessary: decontaminate and
clean surfaces of Autoload ribbon
underneath.
______________________________________
Clean Flat Band Cables Groove
and ensure collision free
movement of Flat Band Cables.
______________________________________
All belts, drives and cables for
any damage, or loose screws.
______________________________________
Recover:
Autoload ribbon
36
______________________________________
Deck Panels
______________________________________
Pipetting Arm
______________________________________
35
Microlab
®
-STAR Instrument with Autoload Option only
36
dito