Maintenance Checklist (continued):
Removal of Trace files
OK REMARKS
Together with the Laboratory
Supervisor check for trace files
witch may deleted from the
Operating PC.
______________________________________
Check for Traces files such as
method / system trace files
communication trace files
______________________________________
Check for old methods which
may be removed from PC
______________________________________
Back them up if necessary, else
delete
______________________________________
Check for available disk space
onto hard drive, 500 MB
minimum are needed
______________________________________
ML STAR IVD additionally:
Test tightness of seal for all
Pipetting Heads.
______________________________________
Check the pressure adjustment
values of TADM.
______________________________________

Table of Contents