Loading...
91
────────────────────────────────────────────────────
7.4 Message Reference
────────────────────────────────────────────────────
Synta
x
:VRANge?
Respons
e
synta
x
(If headers are ON)
:VRANGE data
(If headers are OFF)
data
Exampl
e
(If headers are ON)
Transmission
:VRAN?
Response
:VRANGE 50E+0
(If headers are OFF)
Transmission
:VRAN?
Response
50E+0
Synta
x
:ZERoclear
Exampl
e
(If headers are ON)
Transmission
:ZER
The zero-clear is executed.
F
unctio
n
This command returns the currentl
y
set voltage range in NR3 numeric
data (data) format.
Note
s
If resistance measurement mode is
active, an execution error occurs.
If auto range is active, returns the
currently set range.
F
unctio
n
Zero-clear is a function used to
return the zero-adjust data to their
default values.
:VRANge?
Queries the voltage range.
:ZERoclear
Excuses zero clear.