69
────────────────────────────────────────────────────
7.4 Message Reference
────────────────────────────────────────────────────
Synta
x
*OPC?
Respons
e
synta
x
(If headers are ON)
*OPC?
(If headers are OFF)
1
Exampl
e
Transmission
RST;:MEAS:RES?;*OPC?
Returns a "1" when all *RST and
MEAS actions are complete.
Synta
x
*RST
F
unctio
n
The same as the *OPC command,
except in that, at the instant that
the previous commands have been
completed, instead of bit 0 (the OPC
bit) of the standard event status
register (SESR) being set to 1, the
response message "1" is returned.
F
unctio
n
Resets the 3560 instrument.
The table below lists contents to be
reset.
*OPC?
Returns a "1" instead of setting the SESR bit0 (OPC bit) to "1."
*RST
Initializes the settings.
Initialized items by the 3560
Measurement mod
e
Resistance
measurement mod
e
Comparator numbe
r
1
Resistance
measurement rang
e
Auto range voltage limiter
ON
Sampling rate SLOW
Broken SENSE line
detector
OFF
Hold Free run Comparator mode
AUTO
Comparator OFF
Power supply
frequency
50 Hz
Uneffected contents
Status byte register
Standard event status register
Enable registers.
Interface function
GP-IB address
Output queue
Input buffer
Current pass
Initialized items by the RS-232C or GP-IB
Header
ON
Comparator number to be set by the "CSET" message
1