76
────────────────────────────────────────────────────
7.4 Message Reference
────────────────────────────────────────────────────
Synta
:CSET:RPARameter data1, data2
dat
Rang
(Ω)
data1 data2
30 m 0 to 31.000E-3 0 to 31.000E-3
300 m 0 to 310.00E-3 0 to 310.00E-3
3 0 to 3.1000 0 to 3.1000
30 0 to 31.000 0 to 31.000
300 0 to 310.00 0 to 310.00
3k 0 to 3.1000E+3 0 to 3.1000E+3
*Numerical value dat
Exampl
Transmissio
:CSET:RRAN 30E‑3;:RPAR 10.123E
3, 25.567E‑3
The comparator function executes
in the 30 mΩ range under the
conditions: the upper limit is
25.567 mΩ and the lower limit is
10.123 mΩ.
Synta
:CSET:RPARameter?
Respons
synta
(If headers are ON)
:CSET:RPARAMETER data1, data2
(If headers are OFF)
data1, data2
Exampl
(If headers are ON)
:CSET:RPARAMETER 25.567E‑
3,10.123E‑3
(If headers are OFF)
unctio
This command sets the resistance
range of the comparator specified by
the :CSET:NUMBer command. The
upper and lower limits are
determined automatically; the
smaller value is defined as the lowe
limit and the larger value is defined
as the upper limit.
Note
If the set range deviates from the
range specified by the :CSET:RRANge
command, an execution error occurs
unctio
This command returns the set
resistance range of the comparator
specified by the :CSET:NUMBer
command in NR3 numeric data
(data) format.
:CSET:RPARameter
■ Sets the upper and lower limits for comparator resistance.
:CSET:RPARameter?