79
────────────────────────────────────────────────────
7.4 Message Reference
────────────────────────────────────────────────────
Synta
x
:CSET:VRANge data
dat
a
-50 to 50 (Numerical value data)
Exampl
e
Transmission
:CSET:VRAN 5
The 5 V range is selected in the
comparator setting.
Synta
x
:CSET:VRANge?
Respons
e
synta
x
(If headers are ON)
:CSET:VRANGE data
(If headers are OFF)
data
Exampl
e
(If headers are ON)
Transmission
:CSET:VRAN?
Response
:CSET:VRANGE 5E+0
(If headers are OFF)
Transmission
:CSET:VRAN?
Response
5E+0
Synta
x
:CTMode data
dat
a
AUTO/MANUal (Character data)
Exampl
e
Transmission
:CTM AUT
The external control terminal
mode is set to AUTO.
F
unctio
n
This command sets the voltage
range of the comparator specified by
the :CSET:NUMBer command.
Note
s
If numeric data is not specified, an
execution error occurs.
If resistance measurement mode ha
s
been selected by the :CSET:MODe
command for the comparator
specified by the :CSET:NUMBer
command, an execution error occur
s.
F
unctio
n
This command returns the set
voltage range of the comparator
specified by the :CSET:NUMBer
command in NR3 numeric data
(data) format.
5 V range 5E+0
50 V range 50E+0
Note
s
If resistance measurement mode ha
s
been selected by the :CSET:MODe
command for the comparator
specified by the :CSET:NUMBer
command, an execution error occurs
.
F
unctio
n
This command sets comparator
mode to either of auto-comparator
mode or manual comparator mode.
:CSET:VRANge
Sets for the voltage range used by the comparator.
:CSET:VRANge?
Queries for the voltage range used by the comparator.
:CTMode
Sets the comparator mode.

Table of Contents