7.4 Message Reference
────────────────────────────────────────────────────
Synta
x
:MEASure:VOLTage?
Respons
e
synta
x
(If headers are ON)
:MEASURE:VOLTAGE data1,data2
(If headers are OFF)
data1,data2
data1
Voltage rang
e
(V)
All of sampling rate
5 ±□.□□□□E+0
50 ±□□.□□□E+0
OFF
±1.0000E+8
NG
1.0000E+9
data2
FAIL/PASScomparator results
OFF Comparator is not used.
NG Abnormal measurements
Exampl
e
(If headers are ON)
Transmission
:MEAS:VOLT?
Response
:MEASURE:VOLTAGE 3.5678E+0,FAIL
(If headers are OFF)
Transmission
:MEAS:VOLT?
Response
3.5678E+0,FAIL
Synta
x
:MODe data
dat
a
R/RV (Character data)
Exampl
e
Transmission
:MOD RV
Measurement mode is set to
resistance and voltage
measurement mode.
F
unctio
n
This command returns the current
measured voltage in NR3 numeric
data format (data1), and the
comparator result in
HI/IN/LO/OFF/NG character data
format (data2)or
FAIL/PASS/OFF/NG character data
format (data2).
Note
s
If resistance measurement mode is
active, an execution error occurs.
F
unctio
n
Sets the measurement mode.
:MEASure:VOLTage?
Queries for the current measured voltage and comparator result.
:MODe
Sets the measurement mode.

Table of Contents