EasyManua.ls Logo

Hioki 3560 - Page 98

Hioki 3560
156 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
86
────────────────────────────────────────────────────
7.4 Message Reference
────────────────────────────────────────────────────
Synta
x
:MEASure:RESistance?
Respons
e
synta
x
(If headers are ON)
:MEASURE:RESISTANCE data1, data2
(If headers are OFF)
data1, data2
HI/IN/LO/FAIL/PASS comparator
results
OFF Comparator is not used.
NG Abnormal measurements
data1 Resistance
Resistance
range
(Ω)
Sampling rate
FAST MEDIUM/SLOW
30 m □□.□□E-3 □□.□□□E-3
300 m □□□.E-3 □□□.□□E-3
3 .□□□E+0 .□□□□E+0
30 □□.□□E+0 □□.□□□E+0
300 □□□.E+0 □□□.□□E+0
3k .□□□E+3 .□□□□E+3
OFF
1.0000E+8 1.0000E+8
NG
1.0000E+9 1.0000E+9
data2
HI/IN/LO/FAIL/PASS comparator
results
OFF Comparator is not used.
NG Abnormal measurements
Exampl
e
(If headers are ON)
Transmission
:MEAS:RES?
Response
:MEASURE:RESISTANCE 20.123E‑3,IN
(If headers are OFF)
Transmission
:MEAS:RES?
Response
20.123E‑3,IN
F
unctio
n
This command returns the current
measured resistance in NR3
numeric data format (data1), and
the comparator result in
HI/IN/LO/OFF/NG character data
format (data2)or
FAIL/PASS/OFF/NG character data
format (data2).
HI/IN/LO is returned in voltage
measurement mode, while
FAIL/PASS is returned in
resistance and voltage
measurement mode.
:MEASure:RESistance?
Queries for the current measured resistance and comparator result.