EasyManua.ls Logo

Hitachi TM3030 Plus - Page 176

Hitachi TM3030 Plus
384 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4 - 57
6c. EDX
Well-suited for observing specimens that do not produce adequate brightness /
contrast levels. Select [EDX] for EDX analysis in addition to general sample
observation.
The [EDX] button operation in [Observation condition settings area] allows the same
process to be performed.
NOTE
: Changing the observation condition settings can adversely affect the beam axis
conditions. In such a case, perform beam axis adjustments.
For the details, refer to 5.2 Beam Axis Alignment.
7. Observation Surface
Figure 4.3.6-6 Setting Menu (Observation Surface Menu)
[5 kV] of the Observation condition can be selected along with [Low mag.], [Middle mag.]
or [High mag.] Similarly, every time the [5 kV] button operation in [Observation condition
settings area] is clicked, [Low mag.], [Middle mag.] and [High mag.] can be switched in
order.
For the details, refer to 4.3.12 Observation Condition Selection Button.
8. Image Optimize
Figure 4.3.6-7 Setting Menu (Image Optimize Menu)
Optimizes an image to be saved when executing image save. This function does not
apply to a live display image.
Image optimize has the following three functions:

Table of Contents

Related product manuals