EasyManua.ls Logo

Hitachi TM3030 Plus - Page 252

Hitachi TM3030 Plus
384 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5 - 15
4c. When the image on the observation screen turns into an image as shown in Figure
5.2-10, click the [STEP 2] tab.
Figure 5.2-10 STEP 1 Reference Image
NOTE
: Figure 5.2-10 is only an example. Don’t be alarmed if the same image does not
appear; it is not a problem.
The following window appears during the setting of conditions. Axial alignment should
not be performed until the window disappears.
Figure 5.2-11 Electron Beam Condition Setup Window (2)

Table of Contents

Related product manuals