Invacare®LiNX
•Forward
•Reverse
•Left
•Right
•Ignored
IfIgnoredisselected,thewheelchairdoesnotmovein
anydirectionwhentheinput’sleftquadrantisselected.
RightQuadrantOperation
Thisparametersetsthedirectionthatthewheelchair
moveswhentherightquadrantoftheuserinputis
selected.Optionsavailableare:
•Forward
•Reverse
•Left
•Right
•Ignored
IfIgnoredisselected,thewheelchairdoesnotmovein
anydirectionwhentheinput’srightquadrantisselected.
10.10EnableIndirectNavigation
Dependingonwhichtypeofindirectnavigationshould
beactivatedforuser,differentmainparametersmustbe
enabled.
Typesofindirectnavigationare:
•MenuSelect
•MenuScan
ParameterMenuScanMenuSelect
NavigationType
MenuScan
MenuSelect
EnableMenu
Scan
On
Off
Navigation
TimeoutEnabled
On/OffOn/Off
EnablingMenuScan/MenuSelect
1.OpenUserPreferences.
2.
Fig.10-34
SetparameterNavigationtypeaccordingtoabove
table.
3.OpenFunctions→desiredprole.
4.
Fig.10-35
SetParameterEnableMenuScanaccordingtoabove
table.
5.Opendesiredfunctionwithinthisprole.
6.
Fig.10-36
SetparameterNavigationTimeoutEnabledtoOffor
On.
7.IfNavigationTimeoutEnabledissettoOff,program
anexternalswitchtoenternavigationmode.See
10.11.1ConguringExternalSwitches,page135
.
OtherParametersinUserPreferences
1321605129-G