IEEE-488 Commands
Copyright © Itech Electronic Co., Ltd. 15
*PSC
This command controls whether a service request will generate when the
power supply is re-generalized.
1 OR ON: when the power supply is generated, the values in the Status
Byte Enable register, Operational Event Enable register, Query Event
Enable register and Standard Event Enable register are cleared.
0 OR OFF: the values in the Status Byte Enable register, Operational Event
Enable register, Query Event Enable register and Standard Event Enable
register are saved in NVM, which will be used at regenerating.
Command syntax
*PSC <bool>
Arguments
0|1|ON|OFF
Query syntax
*PSC?
Returns
0|1
*RCL
This command restores the power supply to a state that was previously stored
in memory with a *SAV command to the specified location. All states are
recalled with the following exceptions:CAL:STATe is set to OFF
The trigger system is set to the Idle state by an implied ABORt command (this
cancels any uncompleted trigger actions)
NOTE: The device state stored in location 0 is automatically recalled at power
turn-on.
Command syntax
*RCL <NRf>
Arguments
0 to 49
Example:
*RCL 3

Table of Contents

Other manuals for ITech IT6400 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ITech IT6400 Series and is the answer not in the manual?

ITech IT6400 Series Specifications

General IconGeneral
BrandITech
ModelIT6400 Series
CategoryPower Supply
LanguageEnglish

Related product manuals