Trigger commands
Copyright © Itech Electronic Co., Ltd. 76
2. Clear the WTG bit in the Status Operation Condition register after both
transient and acquire trigger sequences have completed. (WTG is the
logical-or of both transient and acquire sequences.)
3. If INITiate:CONTinuous ON has been programmed, the trigger subsystem
is immediately re-enabled for subsequent triggers. As soon as it is cleared,
the WTG bit is again set to 1.
Command syntax
TRIGger[:SEQuence1][:IMMediate]
TRIGger[:IMMediate]
Arguments
None
Example
TRIG
TRIG:IMM
Related Command
ABOR CURR:TRIG INIT *TRG VOLT:TRIG
TRIGger:SOURce
This command selects the trigger source for the output transient trigger system.
Since BUS is the only trigger source for transient triggers, this command does
not need to be used. It is included for completeness only.
Command syntax
TRIGger[:SEQuence1]:SOURce<source>
TRIGger[:TRANsient]:SOURce<source>
Arguments
BUS
Example
TRIG:SEQ:SOUR BUS
Query syntax
TRIGger:SEQuence:SOURce?

Table of Contents

Other manuals for ITech IT6400 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ITech IT6400 Series and is the answer not in the manual?

ITech IT6400 Series Specifications

General IconGeneral
BrandITech
ModelIT6400 Series
CategoryPower Supply
LanguageEnglish

Related product manuals