Trigger commands
Copyright © Itech Electronic Co., Ltd. 83
TRIGger:SEQuence2:SOURce
TRIGger:ACQuire:SOURce
These commands select the trigger source for measurement triggers as
follows:
BUS GPIB device, *TRG, or <GET> (Group Execute Trigger)
INTernal trigger is generated internally when the measured waveform
crosses the trigger level with the selected slope.
EXTernal Selects the external trigger input
Command syntax
TRIGger:SEQuence2:SOURce<source>
TRIGger:ACQuire:SOURce<source>
Arguments
BUS | INTernal|EXTernal
RSTValue
INTernal
Example
TRIG:ACQ:SOUR BUS
Query syntax
TRIGger:SEQuence2:SOURce?
TRIGger:ACQuire:SOURce?
Returns
<CRD>
TRIGger:SEQuence:MODE
TRIGger:ACQuire:MODE
This command sets the internal voltage trigger mode. Parameters include
AUTo|NORMal|SINGel, i.e., automatic, normal and single modes.
Command syntax
TRIGger:SEQuence:MODE