Measurement Commands
Copyright © Itech Electronic Co., Ltd. 51
Related Command
SENS:SWE:POIN SENS:SWE:OFFS MEAS:ARR
SENSe:SWEep:FREQuency
This command sets the sampling interval, i.e., sampling rate. In general, the set
value can be exactly divided by 30000, as the maximum sampling rate of the
equipment is 30K.
Command syntax
SENSe:SWEep:FREQuency <NRf+>
Arguments
1-30000
RST Value
30000
Example
SENS:SWE:FREQ 15000
Query syntax
SENSe:SWEep:FREQuency?
Returns
<NR3>
Related Command
SENS:SWE:POIN SENS:SWE:OFFS MEAS:ARR
SENSe:WINDow
This command sets the window function that is used in dc and in ac+dc rms
measurement calculations. The following functions can be selected: HANNing |
RECTangular|AVERaging.
Command syntax
SENSe:WINDow[:TYPE] <type>
Arguments
HANNing | RECTangular|AVERaging