Output command
Copyright © Itech Electronic Co., Ltd. 59
Example
OUTP:PROT:DEL 75E-1
Query syntax
OUTPut:PROTection:DELay?
Returns
<NR3>
Related Command
OUTP:PROT:CLE CURR *RCL *SAV
OUTPut:RI:MODE
This command selects the mode of operation of the Remote Inhibit protection.
The RI mode is stored in non-volatile memory. The following modes can be
selected:
Latch: Latch mode. After the detection of level change (from high to low) at the
external On/Off control terminals, the power supply output will be switched off.
To release the latch state, send the command (OUTPut: PROTection: CLEar)
or select “Menu > System > Sys Conf” on the interface to enter the System
Menu interface. Select the System Protection Clear by the Up/Down key. Press
the right [Enter] soft key to release the latch mode.
Live: Live mode. The output state of power supply changes with the level of
external On/Off control terminal. When the On/Off control terminal input is
high-level, the power supply output is on; and when the external On/Off control
terminal input is low-level, the power supply output is off.
Command syntax
OUTPut:RI:MODE <mode>
Arguments
LATCh | LIVE
Example
OUTP:RI:MODE LIVE
Query syntax
OUTPut:RI:MODE?

Table of Contents

Other manuals for ITech IT6400 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ITech IT6400 Series and is the answer not in the manual?

ITech IT6400 Series Specifications

General IconGeneral
BrandITech
ModelIT6400 Series
CategoryPower Supply
LanguageEnglish

Related product manuals