Output command
Copyright © Itech Electronic Co., Ltd. 58
Related Command
OUTP:DFI
OUTPut:PROTection:CLEar
This command clears the latch that disables the output when an overvoltage,
overcurrent, overtemperature, or remote inhibit status condition is detected. All
conditions that generate the fault must be removed before the latch can be
cleared. The output is then restored to the state it was in before the fault
condition occurred.
Command syntax
OUTPut:PROTection:CLEar
Arguments
None
Example
OUTP:PROT:CLE
Related Command
OUTP:PROT:DEL *RCL *SAV
OUTPut:PROTection:DELay
This command sets OCP protection delay, which prevents the subsystem
status bit from being recompiled and causing mis-operation due to
transient-varied CC status.
Command syntax
OUTPut:PROTection:DELay <NRf+>
Arguments
0 to 60
Unit
seconds
RST Value
0.08 (Normal)