Measurement Commands
Copyright © Itech Electronic Co., Ltd. 44
Returns
<NR3>
Related Command
MEAS:CURR?
MEASure:VOLTage:ACDC?
FETCh:VOLTage:ACDC?
These queries return the ac+dc rms output voltage.
Command syntax
MEASure[:SCALar]: VOLTage :ACDC?
FETCh[:SCALar]: VOLTage :ACDC?
Arguments
None
Example
MEAS: VOLT:ACDC?
FETC: VOLT:ACDC?
Returns
<NR3>
Related Command
MEAS:CURR:ACDC?
MEASure:VOLTage:HIGH?
FETCh:VOLTage:HIGH?
These queries return the High level voltage of a voltage pulse waveform. The
instrument first measures the minimum and maximum data points of the pulse
waveform. It then generates a histogram of the pulse waveform using 16 bins
between the maximum and minimum data points. The bin containing the most
data points above the 50% point is the high bin. The average of all the data
points in the high bin is returned as the High level. If no high bin contains more
than 1.25% of the total number of acquired points, then the maximum value is