Measurement Commands
Copyright © Itech Electronic Co., Ltd. 38
Returns
<NR3> [,<NR3>]
Related Command
SENS:SWE:TINT SENS:SWE:OFFS
MEASure:ARRay:VOLTage?
FETCh:ARRay:VOLTage?
These queries return an array containing the instantaneous output voltage in
volts. The output voltage or current is digitized whenever a measure command
is sent or an acquire trigger occurs. The time interval is set by
SENSe:SWEep:TINTerval. The position of the trigger relative to the beginning
of the data buffer is determined by SENSe:SWEep:OFFSet. The number of
points returned is set by SENSe:SWEep:POINts.
Command syntax
MEASure:ARRay:VOLTage?
FETCh:ARRay:VOLTage?
Arguments
None
Example
MEAS:ARR:VOLT?
FETC:ARR:VOLT?
Returns
<NR3> [,<NR3>]
Related Command
SENS:SWE:TINT SENS:SWE:OFFS
MEASure:CURRent?
FETCh:CURRent?
These queries return the dc output current.