IEEE-488 Reference
4-52
:CONFigure:<function>
where; <function> = VOLTage[:DC] DCV function
= CURRent[:DC] DCI function
= VOLTage:AC ACV function
= CURRent:AC ACI function
= RESistance
Ω
2 function
= FRESistance
Ω
4 function
= TEMPerature TEMP function
= FREQuency FREQ function
= VOLTage:FREQuency FREQ function (voltage input)
= CURRent:FREQuency FREQ function (current input)
Formats
:conf:volt[:dc]
:conf:curr[:dc]
:conf:volt:ac
:conf:curr:ac
:conf:res
:conf:fres
:conf:temp
:conf:freq
:conf:volt:freq
:conf:curr:freq
Query
:CONFigure? Query the selected function
Short-form format: :conf?
Response message: CURR:DC, VOLT:AC, CURR:AC, VOLT:DC, RES, FRES, TEMP,
FREQ, VOLT:FREQ or CURR:FREQ
Description
This command is used to configure the instrument for subsequent measurements on the specified
function. Basically, this command places the instrument in a “one-shot” measurement mode.
The :READ? command can then be used to trigger a measurement and acquire a reading (see
:READ?).
Note that there are three <function> versions to configure frequency; FREQuency, VOLT-
age:FREQuency and CURRent:FREQuency. When FREQuency is sent, the *RST input source
is selected (voltage). When :VOLTage:FREQuency is sent, voltage input is selected. When
CURRent:FREQuency is sent, current input is selected.
When this command is sent, the Model 2001 will be configured as follows:
• The function specified by this command is selected.
• All controls related to the selected function are defaulted to the *RST values.
• Continuous initiation is disabled (:INITiate:CONTinuous OFF).
• The control sources in all three layers of the Trigger Model are set to Immediate.
• The count values in all three layers of the Trigger Model are set to one (1).
• The delay in all three layers of the Trigger Model are set to zero (0).
• The Model 2001 is placed in the idle state.